Short Description: CellXpert C170i is a reliable 170 L CO2/O2 incubator that provides a stable, uniform environment for mammalian cell culture by precisely controlling temperature, CO2/O2 levels with fast recovery and low gas consumption. It features high-temperature sterilisation, intuitive touchscreen control, contamination prevention for cell culture application.
Produser: Eppendorf
Equipment Information:
Functions:
- 170 L CO2 incubator for stable mammalian cell culture with precise temperature control (+4 °C–50 °C) and CO2 regulation (0.1–20%).
- Optional O2 control (hypoxic range) for physiologic and hypoxia research applications.
- Fan-less, convection-based airflow system for uniform conditions and reduced contamination risk.
- Dual-beam IR CO₂ sensor for accurate measurement and rapid recovery (5 min) after door opening.
- Certified 180 °C high-temperature sterilisation cycle for reliable chamber decontamination.
- Seamless stainless-steel chamber with rounded corners for easy cleaning and GMP-compliant workflows.
- Removable water tray and high-humidity environment to minimise evaporation.
- Flexible configuration with 4 standard shelves and inner glass doors to reduce parameter fluctuation.
- Low gas consumption with intelligent gas management to reduce operating costs.
- Continuous monitoring of temperature, CO2, and optional O2 levels for process security.
- Designed for stem cells, primary cells, cancer research, and standard tissue culture applications.
Operating Principle:
- Precise thermal regulation combined with infrared gas sensing and controlled humidified convection to maintain a stable, contamination-controlled cell culture environment.
Technical Specification:
- Capacity: 170 L
- Shelves: 4 with individual inner doors
- Maximum load per shelf: 10 kg
- CO2 range: 0.1–20%
- O2 range: 1–20%
VU ID: 8151874
Location: BVI 121
Core Facility: Eukaryotic Cell Engineering
Equipment is operated by Personnel only: No
It is allowed to work with the equipment independently after familiarization with the work process on site: Yes
Only those with the appropriate certificate/certificate are allowed to work with the equipment: No
There is an opportunity to participate in equipment usage training: No
Contact Scientist (Leading Scientist):Indrė Dalgėdienė
Contact email address:
Information for the User:
- Currently, there is no possibility to use O2 gas.
- CO2 Incubators Operating manual CellXpert C170i